Testing, Certification & Repair specialists
of controlled environment equipment

SEPS Team
We are here to serve you

Chris Gorman
CEO/ OPERATIONS MANAGER
cgorman@sepsservices.com


Jayne Strippoli
OFFICE MANAGER
jstrippoli@sepsservices.com



SALES
sepsoffice@sepsservices.com


ceta logo

Certifications

BIOSAFETY CABINET CERTIFICATION AND MAINTENANCE - BSC

SEPS offers comprehensive testing, certification and maintenance of biosafety cabinets. All biological safety cabinets (BSC's) are certified by NSF accredited technicians to meet NSF-49 F guidelines and manufacturer specification. SEPS technicians follow strict SEPS standard operating procedures (SOP) which remain in compliance with Federal and International Safety Guidelines.

Tests Performed in all certification

Inflow velocity test

Provides personal protection

Down flow velocity test

Provides product protection

Hepa filter leak test

Provides protection to your working agents and room environment

Smoke visualization test

Verifies proper airflow direction

Site installation test & alarm calibration

Required on B1 and B2 biosafety cabinets only

Light intensity test

Assures sufficient light to prevent operator fatigue

UV light test

Assures UV bulb is providing proper sterilization

Electrical safety test

Assures that the unit is electrically safe for worker use

Sound test

Verifies that the use of the biosafety cabinet will not contribute to workers fatigue


Optional tests

Vibration testing

Assures that the product is not damaged by vibration

Viable and non-viable particle count

See clean room section for more detail

Pressure decay and cabinet integrity test

Assures no contamination escaping through welds and joints

FUME HOODS - FH

Chemical fume hoods are used to contain and exhaust chemical vapors. They come in different types and designs for different operations. Airflows in fume hoods are imperative for safe operation. SEPS offers airflow velocity tests and smoke patterns visualization tests as means of documenting a fume hood performance. Periodic testing, preventative maintenance and inspections can provide your facility a complete program for fume hood management.

ANSI/ASRAE 110

Method of testing performance of laboratory fume hoods. There are different test procedures incorporated into ASRAE 110 test:

  1. Inflow Velocity Test with proper grid.
  2. Smoke Patters Visualization Test.
  3. Tracer Gas Containment Test

ANIMAL RESEARCH PRODUCTS

SEPS offers testing of animal transfer stations, waste disposal units, necropsy units and clean benches. Testing of these units ensures our customers of optimum productivity, versatility and safety of your research animals. Transfer stations are designed to provide animal protection and particulate control during routine cage-changing operations in the laboratory.

Periodic certification is necessary and will provide proper operation and safety for your lab animals. Contact SEPS to schedule a visit risk assessment and offer proper certifications guidelines.

SEPS also certifies and repairs:

  1. Animal Racks.
  2. Containment Racks.
  3. Other types of animal containment units.

LAMINAR FLOW UNITS - LFU

A laminar flow clean bench provides a working environment that has its own filtered air supply. In many cases clean benches or LFU'S provide ISO class 5 environments. Virtually particulate free work area for conducting procedures.

Laminar flow units prevent any contaminated room air from entering the work area and contaminating samples.

Laminar flow units do no protect users from exposure and should never be used when working with any infectious material or volatile toxic chemicals.

SEPS certified LFU's to (IEST) Institute of Environmental Sciences and Technology guidelines and recommended practices. IEST-RP-CC-002 is the current version.